Model Number LTF-VP
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated by olympus.In addition to the findings in b5, evaluation found the white glue at the distal end body joint was missing, the bending section cover was cracked, the light guide lens was cracked, and the video connector case and printed circuit board were cracked.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The foggy image was likely due to water invasion; however, a definitive root cause of the foggy image could not be determined.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the instructions for use (ifu).If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported the laparo-thoraco videoscope had poor quality image.The issue occurred during a procedure.The customer noted there may be a possible internal issue and there was condensation forming on the inside of the tip of the scope.There was no reported patient harm or impact due to this event.During device evaluation at olympus, it was found the scope image failed a fog test.The report is being submitted due the defect found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA LAPARO-THORACO VIDEOSCOPE
Type of Device
LAPARO-THORACO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19211959
MDR Text Key341375713
Report Number9610595-2024-08885
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170382673
UDI-Public04953170382673
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K955403
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-VP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-