• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2024
Event Type  malfunction  
Manufacturer Narrative
The distributor performed a checkout of the equipment and confirmed the reported complaint.The pcb - power management board was replaced to resolve the issue.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
 
Event Description
It was reported that there was a malfunction resulting in power failure.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 620
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key19212873
MDR Text Key341384315
Report Number9710602-2024-00611
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682103985
UDI-Public01008406821039859
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/19/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-