The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 activation feature was very sensitive and light to the touch.It did not take much to activate the hp.They also noted that in order to turn off the activation, they needed to manually move the toggle to the open position.The energy did not automatically stop after they removed their thumb from the toggle.They had the extension cable changed out but it still did not work properly.They turned the power to the power supply off and on and reconnected the cables and still the same.They then opened a new kit and replaced the hp device and the problem stopped.The new hp device worked as expected.The case was completed without any other issues.There was no delay or patient harm reported.
|