Model Number 71992-01-0007 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.Visual inspection has been performed on sensor and no issues were observed.The sensor plug was seated in mount properly.Data was successfully extracted from the returned sensor using approved software.Removed the sensor plug and inspected the plug assembly, no failure mode observed.Current was applied to the sensor to perform accuracy testing while in the test fixture.All results were within specification.The current was applied to the sensor to perform linearity testing while in the test fixture.All results were within specification.Sensor was reprogrammed and simvivo test performed.All results were within specification.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.Therefore, the issue is not confirmed.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
|
|
Search Alerts/Recalls
|
|