• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported the insufflator flow rate display is partly missing due to the front panel failing to light up.The issue occurred during preparation for use.There were no reports of patient harm.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection.And the reportable malfunction was confirmed.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.It has been over 8 years, since the subject device was manufactured.Based on the results of the investigation, flow rate display was partly missing, due to defective lighting caused by failure of the front panel led (light-emitting diode).However, olympus could not identify a definitive root cause of the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19215741
MDR Text Key341494049
Report Number3002808148-2024-04022
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2024
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/03/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-