Model Number 0998-00-0800-53
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that prior to use when checking the unit, the cardiosave intra-aortic balloon pump (iabp) unit shut down.There was no patient involved.
 
Manufacturer Narrative
A getinge field service engineer (fse) spoke with the customer via phone.It was determined that the unit was not properly seated in the cart which prevented the unit from charging.The fse advised the customer how to reinsert the pump into the cart.Once the unit was in the cart, it was left for several hours to charge.The iabp was returned to the cath lab.H3 other text : device not returned to manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key19216926
MDR Text Key341461105
Report Number2249723-2024-01787
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2024
Date Device Manufactured06/30/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-