BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problems Detachment Of Device Component (1104); Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/11/2013
Event Type  malfunction  
Event Description
It was reported that a care giver (cg) reconnected a solution bag that disconnected from the cassette and fell.The event occurred during dwell one on the home choice (hc).The cg cycled power and the hc went to ¿press go to start¿.The home patient was connected while the hc primed again.The technical service representative (tsr) advised the home patient (hp) to start over with new supplies.No patient injury or medical intervention was indicated as a result of this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).As the cassette was not returned and the lot number is unknown, a device analysis cannot be completed.This complaint is for a report of a use error - reuse of single-use product during dwell one where the caregiver reconnected a bag that become disconnected.Use errors and proper user instructions are addressed in ¿the homechoice and homechoice pro apd systems patient at-home guide¿ , which is shipped with every homechoice device.The patient guide instructs the user not to attempt to reuse any disposable supplies.The patient warns the user not to replace empty solution bags or reconnect disconnected solution bags during therapy.The patient warns the user that possible contamination of the fluid or fluid pathways can result if disposables are reused.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3550423
MDR Text Key4216969
Report Number1416980-2014-00265
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K923065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/11/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberL5C4531
Was Device Available for Evaluation? No
Date Manufacturer Received12/11/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age76 YR
-
-