• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS CORK LRG TAP PRI MOD NCK 0DEG 30MM; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS CORK LRG TAP PRI MOD NCK 0DEG 30MM; IMPLANT Back to Search Results
Catalog Number NLS-300000B
Device Problem Noise, Audible (3273)
Patient Problems Inflammation (1932); Pain (1994)
Event Date 05/18/2011
Event Type  Injury  
Event Description
It was reported that: patient is experiencing pain in his left hip along with inflammation.Patient states that the pain started a month ago along with hearing a grinding and squeaking noise.Patient is reporting that he has seen the surgeon, has had the blood work, mri and x-rays done and is awaiting results.
 
Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock met specifications.The complaint history review indicated that there have been similar events for the reported family.The event was confirmed.A voluntary recall ra2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported event is considered to be under the scope of this recall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LRG TAP PRI MOD NCK 0DEG 30MM
Type of Device
IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS CORK
carrigtwohill, cork
EI 
Manufacturer (Section G)
STRYKER ORTHOPAEDICS CORK
ida industrial estate
carrigtwohill, cork
EI  
Manufacturer Contact
keyla navedo
325 corporate dr.
mahwah, NJ 07430
2018315000
MDR Report Key3600508
MDR Text Key4094210
Report Number9616680-2014-00095
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 12/18/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2015
Device Catalogue NumberNLS-300000B
Device Lot Number34095901
Other Device ID Number1010PCM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/18/2013
Initial Date FDA Received01/13/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/22/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberRA2012-067
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
Patient Weight109
-
-