Device Problem Overheating of Device (1437)
Patient Problem Burn, Thermal (2530)
Event Type  Injury  
Event Description
Please refer to importer report # (b)(4).
 
Manufacturer Narrative
The analysis did show that the bearings have been gritty and the handpiece ran out of specification.The heat up was reproducible during a test run.Due to this condition there is a higher friction during the rotation and hence the bearings and therefore the head is heating up.The condition of the bearings is a result of the wear process which takes place during the use.The user instruction requires a visual and functional test prior to each treatment to ensure that the handpiece is running within specification.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLE LUX CONTRA ANGLE 25LPA
Type of Device
DENTAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bahnhofstrasse 20
warthausen 8844 7
GM  88447
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach 88400
GM   88400
351561356
MDR Report Key3610680
MDR Text Key4091481
Report Number3003637274-2014-00001
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K073478
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Other
Initial Date Manufacturer Received 12/13/2013
Initial Date FDA Received01/13/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2006
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-