Model Number 20
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 01/13/2014
Event Type  malfunction  
Event Description
The customer reported that the device was displaying the service indicator.The device had logged event code 9c19.The event code is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no report of patient use associated with the reported issue.
 
Manufacturer Narrative
Corrected information: catalog number, in the initial medwatch report indicates: 3202458.Catalog number, in the initial medwatch report should indicate: 3202487.Supplemental information: physio-control examined the removed therapy pcb assembly and determined that the cause of the reported failure was a diode, designator cr30.The diode was shorted from pins 5 to 9.
 
Manufacturer Narrative
(b)(4): physio-control examined the customer's device and verified the reported failure.Physio replaced the therapy pcb assembly and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key3613054
MDR Text Key4094497
Report Number3015876-2014-00106
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2014
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Date Manufacturer Received03/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-