MEDTRONIC MED REL MEDTRONIC PUERTO RICO PROTECTA XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number D314TRG
Device Problem Pacing Problem (1439)
Patient Problem Weakness (2145)
Event Type  Injury  
Event Description
The patient¿s spouse reported feeling weak when doing anything.A pulmonologist noted that the patient¿s heart rate did not change when walking.The device rate was raised from 30 to 50 paces per minute.The patient was seen by the cardiologist and no issues were noted with the device and no further changes were made.The device remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant products: 6947-58, implantable tachy lead, (b)(6) 2012.A imu49b-52, implantable pacing lead, (b)(6) 2000.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3617236
MDR Text Key4154803
Report Number3004209178-2014-02460
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial
Report Date 11/12/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2013
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/12/2013
Date Device Manufactured09/20/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4592 COMPETITOR IMPLANTABLE PACING LEAD
Patient Outcome(s) Required Intervention;
Patient Age00080 YR
-
-