Model Number M001509100 |
Device Problems
Difficult to Insert (1316); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/16/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that a foreign substance was noted at the tip of the wire.The target lesion was located at an unspecified artery.A 16.016 fathom wire was selected for a embolotherapy treatment procedure.During preparation, it was noted that a silicone substance was seen at the tip of the device thus making the wire unable to be loaded in the renegade catheter.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Event Description
|
It was reported that a foreign substance was noted at the tip of the wire.The target lesion was located at an unspecified artery.A 16.016 fathom wire was selected for a embolotherapy treatment procedure.During preparation, it was noted that a silicone substance was seen at the tip of the device thus making the wire unable to be loaded in the renegade catheter.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Device evaluated by mfr.The device was returned for analysis.There was no damage to the guidewire.Microscopic inspection of the tip did not reveal any irregularities or damage.The outer diameter of the guidewire was measured with a calibrated snap gage; the maximum outer diameter of the guidewire was.0160", which meets specification.Because the renegade used in the clinical event was not received with the fathom guidewire a rubicon 18 catheter was used for functional testing.The fathom guidewire was advanced through the rubicon with no unusual resistance.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Search Alerts/Recalls
|