Model Number 9-ASD-028
Device Problems Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
Patient Problem Embolism (1829)
Event Date 01/17/2014
Event Type  Injury  
Event Description
The 28mm amplatzer septal occluder (aso) was noted to have embolized the following day and was surgically retrieved and the defect was repaired.
 
Manufacturer Narrative
The 28mm aso was received at sjm and decontaminated.The aso was grossly and microscopically examined and no anomalies were found.The aso met dimensional specifications when measured with a calibrated caliper.The aso was loaded into a test 10f loader, deployed and retracted without difficulty or deformation, under non-physiological conditions.The device history record for this product was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with sjm specifications and procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key3625993
MDR Text Key4176446
Report Number2135147-2014-00011
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/06/2016
Device Model Number9-ASD-028
Device Catalogue Number9-ASD-028
Device Lot Number1109071445
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-