ST. JUDE MEDICAL - NEUROMODULATION EON MINI Back to Search Results
Model Number 3788
Device Problem Failure to Shut Off (2939)
Patient Problem Shaking/Tremors (2515)
Event Date 12/19/2013
Event Type  Injury  
Event Description
It was reported the pt felt stimulation when the ipg was turned off.The patient had fallen recently and it was reported the pt was experiencing tremors.The physician undertook surgical intervention and performed thorough intraoperative testing.It was reported the ipg was disconnected several times and the pt detected stimulation even when the ipg turned off.The physician opted to replace the ipg after the tasting.It was reported the pt was experiencing residual stimulation postoperative.Reprogramming resolved the issue.
 
Manufacturer Narrative
Sjm has limited info related to the pt's medical history and is unable to form an opinion as to the relevancy of the pt's history to the event reported.Sjm defers to the pt's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
jennifer st. clair
6901 preston rd.
plano, TX 75024
9725264677
MDR Report Key3629008
MDR Text Key4208463
Report Number1627487-2014-04040
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 01/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2015
Device Model Number3788
Device Lot Number4003182
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/02/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS LEAD: MODEL 3228; IMPLANTED:
Patient Outcome(s) Other;
Patient Age74 YR
-
-