• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4474
Device Problem Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that when a homechoice is ¿plugged into a 220 ac there is a spark noise from the switch¿ (unspecified).It is unknown if there was any patient (pt) involvement.No further information was provided.Additional information was requested but is not available.
 
Manufacturer Narrative
(b)(4).The device is expected to be received and evaluated.An event history log review will be performed.Should any relevant information be obtained from this review that is related to the reported event or through other means, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The evaluation of the device was completed.A capa was initiated to investigate the event further.If additional relevant information is obtained, then a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.An event history review did not confirm the reported condition.Functional and electrical testing was performed with no issues noted.Upon powering up the device, the power entry module was found to produce a spark noise, confirming the reported condition.The cause of the condition was determined to be due to a faulty power entry module.The power entry module was replaced to correct the issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3630716
MDR Text Key3985379
Report Number1416980-2014-05643
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeGR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4474
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/29/2014
Initial Date FDA Received02/17/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received05/12/2014
06/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-