Catalog Number 5C4474 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported that when a homechoice is ¿plugged into a 220 ac there is a spark noise from the switch¿ (unspecified).It is unknown if there was any patient (pt) involvement.No further information was provided.Additional information was requested but is not available.
|
|
Manufacturer Narrative
|
(b)(4).The device is expected to be received and evaluated.An event history log review will be performed.Should any relevant information be obtained from this review that is related to the reported event or through other means, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).The evaluation of the device was completed.A capa was initiated to investigate the event further.If additional relevant information is obtained, then a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.An event history review did not confirm the reported condition.Functional and electrical testing was performed with no issues noted.Upon powering up the device, the power entry module was found to produce a spark noise, confirming the reported condition.The cause of the condition was determined to be due to a faulty power entry module.The power entry module was replaced to correct the issue.Should additional relevant information become available, a supplemental report will be submitted.
|
|