Catalog Number 5100088000 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/16/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the tps micro driver kept rotating after release of the trigger during a procedure.The case was completed successfully without any procedure delay.There was no medical treatment or intervention required as a result of the event and no adverse consequences.
|
|
Event Description
|
It was reported that the tps micro driver kept rotating after release of the trigger during a procedure.The case was completed successfully without any procedure delay.There was no medical treatment or intervention required as a result of the event and no adverse consequences.
|
|
Manufacturer Narrative
|
Failure analysis is ongoing; additional information may be submitted once the results analysis is complete.
|
|
Manufacturer Narrative
|
The reported event was not duplicated; however, corroded electrical sockets were found by the diagnostic engineer during the device inspection.The presence of corroded sockets may cause or contribute to the reported event.
|
|