Model Number 3716
Device Problems Delayed Charge Time (2586); Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 01/16/2014
Event Type  Injury  
Event Description
Device 1 of 2.Reference mfr.Report: 1627487-2014-03088.It was reported the pt experienced warming at her scs ipg pocket site while charging which resulted in the inability to fully charge the scs ipg.It was also reported the pt experienced increased recharge burden.The pt was advised to charge the ipg more often for less time.No additional information is available at this time.On 08/01/2012, st.Jude medical, neuromodulation division, sent field action letters to pts related to heating while charging and raised awareness of this issue to pts.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Manufacturer Narrative
This ipg serial number was included in a field correction.Sjm has limited information related to the pt's medical history and is unable to form an opinion as to the relevancy of the pt's history to the event reported.Sjm defers to the pt's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
rachel blackwell
6901 preston road
plano, TX 75024
9725264637
MDR Report Key3632113
MDR Text Key4208005
Report Number1627487-2014-03087
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Remedial Action Other
Type of Report Initial
Report Date 01/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2011
Device Model Number3716
Device Lot Number2835826
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS ANCHOR: MODEL: 1194 (2); IMPLANT DATE:; SCS LEAD: MODEL: 3186 (2)
Patient Outcome(s) Other;
Patient Age75 YR
-
-