• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-022
Device Problems Component Falling (1105); Unstable (1667)
Patient Problem Obstruction/Occlusion (2422)
Event Date 01/30/2014
Event Type  Injury  
Event Description
A 22mm amplatzer septal occluder (aso) was implanted; however, a trans-esophageal echo revealed the aso was unstable within the defect.As the 22mm aso was attempted to be percutaneously snared, it fell off the snare catheter and embolized into a pulmonary artery.The 22mm aso was percutaneously retrieved and removed from the patient and a 24mm aso was successfully implanted.No adverse patient effects were reported.
 
Manufacturer Narrative
The results of this investigation are inconclusive because the product was not returned for analysis; however, review of the device history record confirmed the product met manufacturing requirements prior to shipment.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key3635155
MDR Text Key3983673
Report Number2135147-2014-00014
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date04/30/2018
Device Model Number9-ASD-022
Device Lot Number4127521
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2014
Initial Date FDA Received02/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient Weight50
-
-