Catalog Number 26-1221
Device Problem Failure to Auto Stop (2938)
Patient Problem No Code Available (3191)
Event Date 01/29/2014
Event Type  Injury  
Event Description
The affiliate reported: the perforator did not disengage.The dura mater was torn.Many cortex vessels have been injured.The height of the cranial vault was normal.The surgery has been prolonged by more than 30 min.
 
Manufacturer Narrative
Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
Upon completion of the investigation, it was noted that the supplier evaluation revealed the following: the customer's complaint "the perforator did not disengage" was not verified.The perforator met functional test method acceptance requirements: proper engagement and disengagement was achieved with every drilled hole, there was no premature disengagement, or erratic and poor cutting action.The device history records for this perforator (lot# jf021s) were reviewed.All tests and inspections associated with the assembly process met specification requirements.Trends will be monitored for this and similar complaints.At the present time this complaint is considered closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key3641520
MDR Text Key4350110
Report Number1226348-2014-11234
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberJF021S
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/31/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1