• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C8320
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that a homechoice pro device did not work.The issue was not specified.Patient involvement was also not specified.No patient injury or medical intervention was reported as a result of this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).A request for the return of the device has been made.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was returned to baxter and an evaluation was performed to investigate the reported event.Functional testing and visual inspections were performed and the device was determined to meet product specification requirements related to the reported event.A review of the event history log was performed and revealed no failures or malfunctions that were the same as or similar to the reported event.During evaluation, the reported problem was confirmed when it was noted that the power supply had to be replaced.The power supply was replaced during servicing.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3647972
MDR Text Key4194257
Report Number1416980-2014-06777
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberR5C8320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received04/17/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-