INTUITIVE SURGICAL,INC. PROGRASP FORCEPS INSTRUMENT; ENDOSCOPIC INSTRUMENT Back to Search Results
Model Number 420093-08
Device Problem Split (2537)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2014
Event Type  malfunction  
Event Description
It was reported that the prograsp forceps instrument cable was noted to be split and derailed after a da vinci prostatectomy procedure.No patient harm, adverse outcome or injury was reported.
 
Manufacturer Narrative
The instrument was returned and evaluated.Per the customer reported complaint, failure analysis found that the one grip open cable was frayed at the distal idler pulley.The frayed strands stuck out at the wrist.Failure analysis also observed that the distal idler pulley exhibited a particular indentation located at the edge where the grip cable shows damage.The evidence was inconclusive but the frayed grip cable and pulley indentation may have been damaged by coming into contact with other instruments/objects.The evidence was inconclusive, but the damage was likely due to mishandling/misuse.Failure analysis also observed that the pitch up cable was frayed at the distal clevis hub.The small frayed strands stuck out at the wrist.Failure analysis also observed that the distal end of the main tube had various scratch marks with light material removal.The scratches were.174 -.211 in length and not aligned with the tube axis.No other damage was found.The endowrist instruments instructions for use (ifu) specifically states: general precautions and warnings handle instruments with care.Avoid mechanical shock or stress that can cause damage to the instruments.The customer reported complaint does not itself constitute a mdr reportable event; however, the damaged cable and/or tube abrasions, found during failure analysis evaluation could likely cause or contribute to an adverse event if the failure mode were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRASP FORCEPS INSTRUMENT
Type of Device
ENDOSCOPIC INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA 9408 6
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 94086
MDR Report Key3648680
MDR Text Key4287261
Report Number2955842-2014-01182
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
PMA/PMN Number
K050369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 01/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number420093-08
Device Lot NumberM11130814 507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2014
Is the Reporter a Health Professional? No
Event Location Hospital
Date Manufacturer Received01/31/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES
Patient Age60 YR
-
-