Evaluation: we suspected that the combined impedance of heartwires and heart tissue located between the electrodes of the pacing wires established a low impedance electrical circuit less than the minimum impedance (200 ohms) specified for the pace 203h.The following graph illustrates the maximum output (in v shown on the y axis) vs.The impedance as seen by the pacemaker.If the user increases the amplitude output is increased beyond this level, the pace 203h's current limiter determines excess currents, terminates stimulation and displays the warning message "short circuit.".
|