The cause of the discordant advia centaur total hcg is unknown.A siemens service engineer (fse) was sent to the customer site for system inspection and checked the reagent and sample probe alignment, acid and base dispense and the wash station and luminometer performance.Instrument was operational.The fse ran quality control material in 20 replicates and all results were within range and showed good precision.The instrument is performing within specification.The initial sample was discarded and is not available for further evaluation and the initial results could not be confirmed.The customer could not confirm if the original tube was labelled correctly.No further investigation is required.The limitations section of the instructions for use states: "test results alone are not diagnoses of medical conditions.For example, low titer elevations of hcg can occur in normal non-pregnant subjects.A physician's diagnosis involves evaluation of the test result in conjunction with, and in the context of, the patient's medical history, physical examination, and other test results--sometimes in consultation with other medical experts.".
|