(b)(4).Physio-control evaluated the device and was able to verify the reported service indication and event codes logged, but was unable to verify the reported reset condition with the device.Upon evaluation, physio observed that the flex cable assembly which connects the user interface pcb to the pcb stack was torn at p21, which is the likely cause of the reported reset condition.Physio replaced the flex cable assembly and observed proper device operation through functional and performance testing and returned the device to the customer for use.
|