The affiliate stated the customer reported elevated total beta human chorionic gonadotrophin (tbhcg) results, for three patients, and elevated quality control (qc) involving the access total bhcg reagent used in conjunction with the unicel dxi 800 access immunoassay system.The customer calibrated a new reagent lot and noted it recovered within the established ranges - qc was also within specification.The customer indicated the elevated tbhcg patient results were not released out of the laboratory.There was no report of patient consequence or change in patient treatment associated with this event.Archive data review indicates that higher recovery does occur with the involved reagent lot across multiple reagent packs.The patients' samples were collected in 100x13 mm becton dickinson (bd) tubes and centrifuged at 3,000 rpm (rotations per minute) for ten minutes.The customer indicated the samples were analyzed approximately five hours apart.There was no report of instrument issues at the time of the event.The instrument was in operation.
|
There is no indication that the alleged access total bhcg device was returned for evaluation.The tbhcg reagent lot in question generated patient results of approximately 15-30% greater than the alternate reagent lot and approximately 10% greater than the alternate lot for qc.The product insert states, "the distinct nature of the biological materials used to produce immunoassay reagents, calibrators, and quality controls (qc) contribute to normal variability between lots.This variability, combined with the very low concentrations of analyte immunoassays are designed to measure, typically yields overall imprecision parameters which are higher than chemistry assays.Coefficient of variations (cvs) for chemistry assays are typically 3-6%, while typical cvs for immunoassays are 8-12%." in conclusion, a definitive cause of the incident could not be determined with the available information.
|