Catalog Number 26-1221
Device Problem Failure to Auto Stop (2938)
Patient Problem No Code Available (3191)
Event Date 02/28/2014
Event Type  Injury  
Event Description
The affiliate reported "it was reported that the perforator did not stop drilling after going through the bone and entered the brain tearing the dura and causing some bleeding." the patient is fine and no long term deficit.
 
Manufacturer Narrative
Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
Upon completion of the investigation, the returned perforator was visually inspected as received, disassembled and cleaned, and then visually and dimensionally inspected.No discrepancies were found.The perforator was reassembled and was functionally tested for cutting and drilling.It was found to meet specification requirements.The reported condition could not be duplicated.Review of the device history record has found no discrepancies.At this time, the complaint is considered to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES, & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key3672729
MDR Text Key16058288
Report Number1226348-2014-11301
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberMF022S
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-