Device history review: review of the device history records indicate devices were manufactured and accepted into final stock.Complaint history review: there have been no other events reported for the reported manufacturing lot.Visual inspection: the implant was received in the carton with the shrink wrap ripped at the top from the user opening the box.Slight compression damage was noted to one of the sides of the carton.A loaners sticker is present on the other side of the carton, applied to the shrink wrap.The outer blister and inner blister were both in tact and fully sealed.No signs of any attempts to open the outer blister pack were noted.The implant was oriented in the upper right corner with the pegs facing upward.A hole was observed in the upper right corner of the outer tyvek lid.The outer blister was then opened and a hole, respective to the implant position and hole in the outer lid, was observed in the inner tyvek lid.The inner tyvek lid was opened, and further damage to the tyvek around the hole was observed.White particles were observed on the surface of the implant.This is assumed to be a result of the grit-blasted internal surfaces and pegs rubbing up against the foam.Dimensional and functional inspections were not performed as the reported event relates to a packaging issue.The event was confirmed.The packaging was confirmed to be assembled per specification.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|