Catalog Number 33873
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Information (3190)
Event Type  Injury  
Event Description
A doctor's office alleged that seven (7) patients had experienced a debonding of restoration after placement with the maxcem elite product.This is the fourth of seven (7) reports.
 
Manufacturer Narrative
Patient specifics with regard to age, gender and weight was not provided by the doctor.The doctor cleaned and re-cemented the crown for the patient using the same product, without further incident.To date, the patient is doing fine.An 'adhesive strength' test of the returned product was evaluated, yielding results within specifications.A dhr review revealed that there were no deviations from the manufacturing process.In addition, no similar complaints were received with regard to this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXCEM ELITE
Type of Device
CEMENT, DENTAL
Manufacturer (Section D)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer (Section G)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer Contact
kerri casino
1717 w collins ave
orange, CA 92867
7145167634
MDR Report Key3675021
MDR Text Key21999821
Report Number2024312-2014-00122
Device Sequence Number1
Product Code EMA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 02/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Device Expiration Date11/01/2014
Device Catalogue Number33873
Device Lot Number4878409
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/12/2014
Initial Date FDA Received03/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-