(b)(4).Method: one complaint mr290 autofeed humidification chamber was returned to fisher & paykel healthcare in (b)(4) for evaluation.The chamber was visually inspected.Results: visual inspection revealed that the feedset tube had been torn in two near the bag spike connection.Sufficient glue was found at the spike and feedset tube connection and the glue bond had held.The surface at the tube break was rough, indicating that the tubing had been physically torn in two, possibly as a result of rough handling.A lot check revealed no similar complaints for lot 131001.Conclusion: the damage appears to be caused by the tube being pulled away from the spike; possibly due to the feedset being caught or put under tension.We have conducted extensive testing of the mr290 chamber, with particular emphasis on feedset breaks.Significantly we have not been able to replicate failure of the feedset tube in any of our testing.The specification for the chamber requires that the feedset tube should have a breaking strain of 30 newtons.During production, pull testing of the feedset strength at both spike and dome end is performed every hour on mr290 chambers from each production line.Additionally all chambers are pressure tested before they leave the production line and any holes or leaks in the feedset are identified during this process.Chambers that fail any of these tests are discarded.This suggests that the damage occurred after the product was released for distribution.The user instructions that accompany the mr290 state the following: "set appropriate ventilator alarms."; "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." (b)(4).
|