ST. JUDE MEDICAL - NEUROMODULATION EON MINI Back to Search Results
Model Number 3788
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Delayed Charge Time (2586); Temperature Problem (3022)
Patient Problems Fall (1848); Therapeutic Effects, Unexpected (2099); Burning Sensation (2146)
Event Date 12/27/2012
Event Type  Injury  
Event Description
Device 1 of 2.Reference mfr report: 1627487-2014-26035.It was reported that soon after implant the patient experienced a burning sensation near the implant site.The ipg would turn on full power and turn itself off and would stimulate the sciatic nerve and cause him to fall to the ground.The patient also experienced an increased recharge burden.The ipg was moved to another location on (b)(6) 2012 but this did not resolve the issues.The patient was explanted.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Manufacturer Narrative
This ipg serial number was included in a field advisory and a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
tanja hoffman
6901 preston rd.
plano, TX 75024
9723098520
MDR Report Key3682953
MDR Text Key17158196
Report Number1627487-2014-26034
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Not Applicable
Remedial Action Other
Type of Report Initial
Report Date 12/31/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/01/2013
Device Model Number3788
Device Lot Number3359052
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/31/2013
Initial Date FDA Received01/10/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-002-R
Patient Sequence Number1
Treatment
IMPLANT:; SCS EXTENSION: MODEL 3383
Patient Outcome(s) Other;
Patient Age45 YR
-
-