Catalog Number 2131000000 |
Device Problems
Sticking (1597); Device Inoperable (1663); Malposition of Device (2616)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/26/2014 |
Event Type
malfunction
|
Event Description
|
It was reported via repair work order that the bed lift was stuck at mid height due to a faulty wiring harness.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Event Description
|
It was reported via repair work order that the bed lift was stuck at mid height due to a faulty wiring harness and base extension cable malfunction.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
It was reported via repair work order that the bed lift was stuck at mid height due to a faulty wiring harness.Follow-up submitted as it was also determined the lift being stuck was due to a base extension cable malfunction.
|
|
Search Alerts/Recalls
|