It was reported that during preparation of the 2.5 x 15 mm xience pro stent delivery system (sds), the physician noticed that there was a slight kink directly behind the blue tip of the tip; therefore,the device was not used.A new 2.5 x 15 mm xience pro stent was then used successfully.A 3.0 x 8 mm xience pro sds was then advanced, but could not be placed.It is unknown what additional steps were taken.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.Returned device analysis of the 2.5 x 15 mm xience pro sds found that the soft tip is torn.
|
(b)(4).The xience pro is currently not commercially available in the us.The product code listed and the pma# listed are d on the predicate device (xience v) and is therefore similar to a device sold in the us.Evaluation summary: the device was returned for analysis.The kink on the tip was not confirmed.There was a tear in the soft tip at the very distal end of the tip.Based on a visual inspection analysis of the returned device, there is no indication of a product deficiency.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the electronic complaint handling database revealed no other similar incidents reported from this lot.Based on the reviewed information, no product deficiency was identified.
|