Model Number 1192
Device Problems Kinked (1339); Device Operates Differently Than Expected (2913)
Patient Problem Inadequate Pain Relief (2388)
Event Date 02/03/2014
Event Type  Injury  
Event Description
Device 2 of 2.Ref mfr.Report #: 1627487-2014-21125.The pt received two scs leads and two scs anchors, each have the same lot number.It was reported the pt was not receiving right-sided stimulation.Diagnostics revealed multiple invalid impedances.X-rays did not provide certainty if the lead had migrated, however; there were multiple loops/ kinks observed at the anchor site.Follow-up identified surgical intervention will be taken to address the issue.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
palka sharma
6901 preston rd.
plano, TX 75024
9725264823
MDR Report Key3697141
MDR Text Key4263954
Report Number1627487-2014-21126
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 02/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2014
Device Model Number1192
Device Lot Number3639525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/03/2014
Initial Date FDA Received02/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS IPG, MODEL 3716
Patient Outcome(s) Other;
Patient Age41 YR
-
-