Model Number T176
Device Problems Improper or Incorrect Procedure or Method (2017); Inadequate User Interface (2958); Device Handling Problem (3265)
Patient Problems Bone Fracture(s) (1870); Injury (2348)
Event Date 03/12/2012
Event Type  Injury  
Event Description
Per email received (b)(6) 2012, customer reports the acetabular reamers did not mill.The teeth were totally worn and the surgeon had to apply too much force.The reamer went through the bone and there was a cement leak possibly going to the peritoneum.Customer reports patient injury and that an unanticipated surgical complication occurred.No additional information was provided.
 
Manufacturer Narrative
Review of the returned parts confirms that the teeth are worn and, in some cases, damaged from years of use.Use of the devices in this state is advised against in the instructions for use that accompanies the parts, which states that all cutting edges should be inspected for damage and wear prior to use.The parts should not have been used in this state.No corrective action will be taken.The parts were lifetime worn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR REAMER 54MM CROSS
Type of Device
ACETABULAR REAMER 54MM CROSS
Manufacturer (Section D)
GREATBATCH MEDICAL SA
l' echelette 7
orvin CH-2 534
SZ  CH-2534
Manufacturer (Section G)
GREATBATCH MEDICAL SA
Manufacturer Contact
jennifer meng
2300 bershire lane north
minneapolis, MN 55441
MDR Report Key3697723
MDR Text Key4251495
Report Number9614497-2014-00007
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 03/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberT176
Device Catalogue Number2014-54
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/04/2012
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/22/2012
Initial Date FDA Received03/06/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-