Model Number 1900
Device Problems Device Alarm System (1012); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/20/2014
Event Type  malfunction  
Event Description
The technician reported the nurse call cable is not able to connect, no nurse call function.The bed is located in the bed shop at the facility.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Manufacturer Narrative
The technician found the communication cable has bent pins.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Check for current leakage at the nurse call system communication connections.Warning: a communication cable must be used for beds that have nurse call.Failure to do could cause pt injury.For beds that have nurse call systems, a communication cable must be connected between the bed and facility communication system.The technician replaced the communication cable to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL CARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
jennifer morris
1069 state route 46 east
batesville, IN 47006
8129313121
MDR Report Key3700888
MDR Text Key4228449
Report Number1824206-2014-00468
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/20/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-