MAKO SURGICAL RESTORIS MULTICOMPARTMENTAL KNEE PROSTHESIS Back to Search Results
Device Problem Fracture (1260)
Patient Problem Pain (1994)
Event Date 02/20/2014
Event Type  Injury  
Event Description
A patient had received a partial knee arthroplasty, which was performed using the robotic arm interactive orthopedic system (rio) and restoris multicompartmental knee implants.The patient experienced some pain at approximately 30 days post-operatively.The pain eventually subsided.At the 90 day follow up, the surgeon noticed that the follow-up x-ray showed an anterior tibial fracture, which had healed.
 
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been conducted by mako surgical.The anterior tibial collapse was the reason for the patient's pain.The fracture healed by the 90 day follow up visit, and the patient's pain had subsided.The surgeon reviewed the x-ray and determined that no medical intervention is required.The rio case session files were also reviewed.All values were found to be within acceptable tolerances, and the data suggest that the rio operated as expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORIS MULTICOMPARTMENTAL KNEE PROSTHESIS
Type of Device
COMPARTMENTAL KNEE PROSTHESIS
Manufacturer (Section D)
MAKO SURGICAL
ft. lauderdale FL
Manufacturer Contact
william tapia
2555 davie rd.
ft. lauderdale, FL 33317
9546280605
MDR Report Key3716166
MDR Text Key4233216
Report Number3005985723-2014-00031
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO)
Patient Age70 YR
Patient Weight68
-
-