Final device investigation found that the device returned was an ether420 and was in good visual condition.Upon evaluation, the remaining 11 usable clips were fired.The first three clips had proper pinch but not proper alignment, they were scissored.The fourth clip came out sideways and unformed.The remaining seven clips did not load properly into the jaws and came out of the device sideways and unformed.After the last clip was fired, the locking mechanism engaged as intended.The device had been sent out with 14 clips.The lot number was not provided, so the device history record could not be reviewed.As each device is test fired prior to release, no conclusion could be made as to what may have caused the reported event.
|