ETHICON ENDO-SURGERY, LLC. HARMONIC ACE ADAPTIVE TISSUE TECHNOLOGY; INSTRUMENT, ULTRASONIC SURGICAL
|
Back to Search Results |
|
Catalog Number HAR36 |
Device Problems
Detachment Of Device Component (1104); Flaked (1246); Melted (1385)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/19/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was returned with the tissue pad damaged, melted and a half portion not returned; in addition, the blade was found scratched.The device was tested with a test handpiece and generator and no alert screens were displayed during the analysis.Probable causes of blade damage are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Probable causes of the tissue pad damage is applying pressure between the instrument blade and tissue pad without having tissue between them.And activating the blade without tissue between the blade and tissue pad to avoid damage to the tissue pad.Both conditions can result in probable damage to the instrument.
|
|
Event Description
|
It was reported that during a laprascopic hysterectomy, the coating on the shaft of the instrument flaked off, but was retrieved.During the same procedure, they also used an har36 and the tissue pad melted and then detached from the instrument, but was also retrieved.The doctor was finished with the procedure at this point and there was no patient consequence reported.
|
|
Manufacturer Narrative
|
(b)(4).When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.
|
|
Search Alerts/Recalls
|
|
|