• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC., CRMD PROMOTE RF CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC., CRMD PROMOTE RF CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number 3213-36
Device Problems Premature Discharge of Battery (1057); No Device Output (1435)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2011
Event Type  Injury  
Event Description
It was reported that the device exhibited a prematurely depleted battery and no output.The device was explanted and replaced.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.Longevity estimations were performed and the device did not meet the expected longevity based on device usage.No high current drain sources were found.The battery was destructively analyzed which found no sources of shorting.The cause of premature battery depletion could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE RF CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view court
sylmar, CA 91342
8184932621
MDR Report Key3730117
MDR Text Key16776217
Report Number2938836-2014-08951
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2008
Device Model Number3213-36
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2011
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2011
Initial Date FDA Received04/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-