• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C8320
Device Problem Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/14/2014
Event Type  malfunction  
Event Description
It was reported that the home choice (hc) gave a beeping noise only.The event occurred after turning on the hc.No patient injury or medical intervention was indicated as a result of this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was returned to baxter and the evaluation is complete.A visual inspection was performed.Functional testing was performed and the digital board was determined to need replacement.The event log could not be reviewed due to the related issue of the digital board replacement.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was reported to be available but has not yet been received.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The cause of the reported beeping noise was determined to be a defective printed circuit board.The printed circuit board was replaced to resolve the condition.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3730236
MDR Text Key4322655
Report Number1416980-2014-11216
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberR5C8320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/29/2014
Initial Date FDA Received04/07/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received05/22/2014
06/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-