• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT ABREVO CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT ABREVO CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number TVTOML
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/17/2014
Event Type  malfunction  
Manufacturer Narrative
Conclusion: the actual device involved in this event was returned for evaluation.The device was visually evaluated.Upon evaluation, one needle was bent and broken at two sites on its middle, not at the tip.Based on the product evaluation performed, it seems that when using the device, the helical passer has been removed and reintroduced in the needle creating the bent on the middle of the plastic needle.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Event Description
It was reported that a patient underwent a sling procedure on (b)(6) 2014.During the procedure, the device had a broken trocar sheath tip upon opening.Another like device was used to complete the procedure with no adverse patient consequences.
 
Manufacturer Narrative
(b)(4).The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT ABREVO CONTINENCE SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key3731268
MDR Text Key11899930
Report Number2210968-2014-04406
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2014
Device Catalogue NumberTVTOML
Device Lot Number3740842
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-