The cause of the biased low d-dimer results is user error.The user incorrectly input the calibrator value during a (b)(6) 2014 calibration.The issue was corrected by recalibration of the assay with the correct calibrator value input.The customer called siemens back and stated the pathologist has reviewed all patients affected and determined that there were no re-admission, no re-tests, no results questioned by treating physicians, and no suspicion of any negative effect on patient outcome/treatment.The instrument is performing within specifications.No further evaluation of the device is required.
|