SYNCARDIA SYSTEMS, INC. SYNCARDIA COMPANION HOSPITAL CART; CIRCULATORY ASSIST DEVICE Back to Search Results
Catalog Number 397003-001
Device Problems Battery Problem (2885); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/21/2014
Event Type  malfunction  
Event Description
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer reported the companion 2 driver had 45 minutes of battery life left, even though the companion 2 driver was docked into the companion hospital cart which was plugged into a/c wall power.He switched the hospital cart power cord to a different wall outlet, and this did not remedy the situation.The plug was then wiggled in the hospital cart and the companion 2 driver went back onto wall power with the batteries charging.The customer then taped the cord to the companion hospital cart to prevent the plug from coming dislodged.The customer also reported that there was no pt impact and the pt was still being supported by the companion 2 driver and companion hospital cart.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION HOSPITAL CART
Type of Device
CIRCULATORY ASSIST DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake rd.
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake rd.
tucson AZ 85713
Manufacturer Contact
carole marcot, esq
1992 e. silverlake rd.
tucson, AZ 85713
5205451234
MDR Report Key3737890
MDR Text Key4278319
Report Number3003761017-2014-00039
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397003-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2014
Initial Date FDA Received03/07/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-