MAUDE Adverse Event Report: HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/10/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from a hill-rom technician stating the nurse call would not work.The bed was located in storage.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The hill-rom technician found the communication cable was partially unplugged and the sidecom board had bent pins where the communication cable attaches.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare bed components are functioning as originally designed.Examine and test the communication junction box.Make sure the sidecom communication system feature works correctly.Examine the communication cable, include the male nad female pins in the plug.A search of the hill-rom maintenance records shoed that hill-rom performed preventative maintenance on this bed in 2012 and 2013.It is unknown if the facility performed any other preventative maintenance.The technician replaced the sidecom board to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key3737925
MDR Text Key4312726
Report Number1824206-2014-00781
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/10/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-