Model Number 1840
Device Problem Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/10/2014
Event Type  malfunction  
Event Description
The account reported the bed would raise the hi/low function by itself.The bed was located in room 206 at the facility.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The technician was told by the account that the hi/low would run from the lowest position to the highest position on its own.The technician found the bed was not equipped with a hand pendant and could not duplicate the self-run issue.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Test the bed function controls for proper operation of the function and momentary operation.Test all lookout controls and individually check for proper operation.The technician found the bed functions as designed and no repair is needed.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE BARIATRIC PLUS BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
jennifer morris
1069 state route 46 east
batesville, IN 47006
8129313121
MDR Report Key3737952
MDR Text Key20109242
Report Number1824206-2014-00774
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 02/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1840
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2014
Initial Date FDA Received03/10/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-