Model Number 6215
Device Problem Deflation Problem (1149)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/14/2014
Event Type  malfunction  
Event Description
It was reported that during a procedure, the physician tested the venogram balloon catheter prior to insertion into the guiding system.After inflating the balloon, the balloon expanded in an irregular fashion and did not deflate completely.A different venogram balloon catheter was utilized.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the u.S.Where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: the balloon catheter was returned and the mechanical operation of the balloon catheter was analyzed.The analyst noted that one side of the balloon inflated less than the other side.No apparent damage was noted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN VENOGRAM BALLOON CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3739917
MDR Text Key4223796
Report Number9612164-2014-00377
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K012225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/04/2015
Device Model Number6215
Device Catalogue Number6215
Device Lot Number61331714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-