It was reported that prior to starting a da vinci surgical procedure, the orange tip of the monopolar curved scissors instrument tip was halfway broken, had a split.The planned surgical procedure was completed and no patient harm, adverse outcome or injury was reported.Nothing reportedly fell into the patient.
|
The monopolar curved scissors instrument was returned and evaluated.Per the customer reported complaint, failure analysis investigation found instrument's tube extension broken and missing a piece measuring approximately.270 x.165 at the proximal clevis interface.The tube extension fractured next to one of the keys that mates with the proximal clevis.Failure analysis concluded that the tube extension breakage may have been due to mishandling.Failure analysis investigation also found that the instrument's tube extension had pad printing removed, measuring approximately.115' x.458.In addition, the tube extension had a dark purple discoloration towards the distal end.The tube extension did not feel different when touching the normal colored area vs.The discolored areas.It is unknown if the discoloration is due to reprocessing or from the pad printing application process.No other damage was found.Failure analysis concluded that the pad printing removal may have been due to mishandling.The endowrist® instruments instructions for use (ifu) specifically states: general precautions and warnings handle instruments with care.Avoid mechanical shock or stress that can cause damage to the instruments.The customer reported complaint does not itself constitute a reportable event; however, the additional finding of the pad printing removal from the tube extension could likely cause or contribute to an adverse event, if the malfunction were to recur.
|