BOSTON SCIENTIFIC - FREMONT (CE) MAESTRO 3000? CARDIAC ABLATION SYSTEM - CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M00421000TCR0 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Nerve Damage (1979)
|
Event Date 03/10/2014 |
Event Type
Injury
|
Event Description
|
Same case as mdr id 2134265-2014-02174.It was reported that a phrenic nerve damage occurred.A maestro 3000 cardiac ablation system pod and controller were selected for treatment.During procedure, a noisy fan was observed from the controller.Furthermore, it was noted that the pod had some loose components inside which caused a rattling sound.To ensure that the phrenic nerve was not in the area of ablation, the physician was pacing from a 4mm blazer ii htd ablation catheter.However, it was noted that an non-bsc stimulator was not pacing so the phrenic nerve was not captured thus the phrenic nerve was damaged causing breathing issues to the patient.The procedure was completed with another of the same device and the patient's breathing function has improved.
|
|
Manufacturer Narrative
|
Age at the time of event: 18 years or older.(b)(4).
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: investigation completed.The device was returned to (b)(4) for evaluation.Device analysis noted that the fan was audibly noisy.However, it did not impact the functionality of the device and was not determined to have caused or contributed to the reported patient injury.The maestro controller passed power-on self-test, rf delivery tests, and all performance criteria of the final test procedure (without removal of the device's top enclosure).Environmental stress testing was also performed at high and low temperature and high and low line voltages with no problem found.The maestro controller also passed testing with maestro pod serial number (b)(4) involved in the same incident, with no errors reported and no problem found.During servicing of the maestro controller at (b)(4), the fan assembly was replaced with a new part due to obsolescence.The maestro controller in this complaint was manufactured prior to the fan assembly change.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is anticipated procedural complications as this event is a known physiological effect of the procedure and is noted within the dfu.(b)(4).
|
|
Event Description
|
Same case as mdr id 2134265-2014-02174.It was reported that a phrenic nerve damage occurred.A maestro 3000 cardiac ablation system pod and controller were selected for treatment.During procedure, a noisy fan was observed from the controller.Furthermore, it was noted that the pod had some loose components inside which caused a rattling sound.To ensure that the phrenic nerve was not in the area of ablation, the physician was pacing from a 4mm blazer ii htd ablation catheter.However, it was noted that an non-bsc stimulator was not pacing so the phrenic nerve was not captured thus the phrenic nerve was damaged causing breathing issues to the patient.The procedure was completed with another of the same device and the patient's breathing function has improved.
|
|
Search Alerts/Recalls
|
|
|