Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Loose or Intermittent Connection (1371); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/25/2014
Event Type  malfunction  
Event Description
It was reported that the neurologist¿s handheld device was not working and was unable to interrogate the patient¿s generator.Follow-up confirmed that the handheld device was functioning and successfully communicated with the patient¿s generator.It was noted during follow-up that the handheld device¿s serial cable was loose and had to be securely held for successful interrogation.
 
Event Description
Additional information was received indicating that there were no issues with the programming system.The handheld device was able to successfully interrogate a demonstration device multiple times.The serial cable was not loose but was provided to the neurologist as a possible cause of communication issues for troubleshooting purposes.The programming system was stored plugged into an outlet and placed in a drawer.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records confirmed that the handheld device passed all functional tests prior to distribution.Serial #, lot #, other, corrected data: previously submitted mdr indicated this information was unknown.It is now known.This report is being submitted to correct this data.Name and address, fax #, corrected data: previously submitted mdr indicated this information was unknown.It is now known.This report is being submitted to correct this data.Manufacture date, corrected data: previously submitted mdr indicated this information was unknown.It is now known.This report is being submitted to correct this data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3763969
MDR Text Key4420544
Report Number1644487-2014-01083
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1075856
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 04/23/2014
Initial Date FDA Received04/22/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age49 YR
-
-