• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, NEUROMODULATION OCTRODE; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Detachment Of Device Component (1104); High impedance (1291); Low impedance (2285)
Patient Problem Inadequate Pain Relief (2388)
Event Date 03/12/2014
Event Type  Injury  
Event Description
Device 1 of 3.Reference mfr.Report: 1627487-2014-26270, 1627487-2014-26275.It was reported the patient is experiencing auto-reducing.Xrays were taken and showed the lead has pulled out of the ipg header.Lead diagnostics showed both low and high impedances.Surgical intervention is pending to address this issue.
 
Manufacturer Narrative
Sjm has limited info related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, NEUROMODULATION
plano TX
Manufacturer Contact
tanja hoffman
6901 preston rd.
plano, TX 75024
9723098520
MDR Report Key3764071
MDR Text Key4487646
Report Number1627487-2014-26269
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 03/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2015
Device Model Number3186
Device Lot Number4180991
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/12/2014
Initial Date FDA Received04/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS IPG, MODEL 3788; IMPLANT DATE:; SCS ANCHOR, MODEL 1192 (2); IMPLANT DATE:
Patient Outcome(s) Other;
Patient Age65 YR
-
-